Hamilton MICROLITER syringes are precision measuring
instruments. For prolonged syringe life and to obtain the maximum benefits of use, a few
helpful tips should be observed.
ASSURED ACCURACY AND PRECISION
- In using a MICROLITER syringe, grasp only the syringe flange
and plunger button. By doing so, variations in liquid measurement due to body heat are
avoided.
- Pump the plunger with the syringe needle immersed in the
liquid to be transferred. This willexpel any trapped air in the needle and syringe.
Minimize the use of a "dry" syringe.
- Every MICROLITER syringe is handcrafted with its own unique
plunger to insure only the highest precision liquid transfers. Plungers are not
interchangeable.
- If the plunger is accidentally withdrawn completely from the
syringe barrel, wipe it carefully with a lint-free tissue before reinserting into the
barrel. Avoid touching the plunger since oil from one's fingers will often interfere with
proper plunger operation.
CLEANING
- The life of your MICROLITER syringe is directly related to
its cleanliness!
- To clean syringes, it is best to use solvents known to be
effective in solvating the sample and preferably that are non-alkaline, non-phosphate and
non-detergent based..
- High, quality grade water and acetone prove to be good
rinses.
- To clean the plunger, remove it from the syringe barrel and
gently wipe with a lint-free tissue. Reinsert the plunger into the barrel and pump
deionized water or acetone through the needle and syringe. Air dry the syringe for
storage.
SOLVENT COMPATIBILITY
- Some solvents may attack and deteriorate the highly
resistive adhesives (cements) used to affix needles in MICROLITER syringes. Avoid
prolonged immersion of the syringe while cleaning. Rinse the syringe thoroughly after use
with deionized water, acetone or another solvent compatible with the sample. Allow to air
dry.